The Delta ICE 2L maintains 2°C to 6°C storage for 48 hours at 25°C ambient temperature, 24 hours at 44°C, and 12 hours at 60°C. Its temperature-controlled aluminum insert and rugged, insulated outer case are designed to hold performance across a range of field conditions.
Yes. The Delta ICE 2L is FDA Listed (Class II, Code KSE) and is an AABB Standards-Compliant Product. It is also CE compliant for EU safety standards as laboratory equipment.
The Delta ICE 2L holds two (2) 450 mL bags of blood in its 2-liter internal storage volume.
The Delta ICE 2L is IP55-rated against dust and water ingress, with insulation protected from drops and punctures, and operates in ambient temperatures from -20°C to 60°C on battery power (0°C to 45°C when AC powered).
The temperature-controlled aluminum insert requires about 8 hours in a freezer set to -8°C to fully freeze, and approximately 15 minutes at room temperature (25°C) to condition and prepare for use. The accessory Delta TempWatch can be used to alert users when a pack is ready to install in the ICE 2L.
|
PCM pack freezer temperature |
Freezing time |
|---|---|
|
-14°C (6.8°F) |
6 hours |
|
-8°C (17.6°F) |
8 hours |
|
-2°C (28.4°F) |
12 hours |
Yes, it records internal temperature every 15 minutes with ±1°C accuracy, storing up to 3 days of data, and reports readings via Bluetooth and cellular connectivity to the BloodCOMM® app.